Nurse-based evaluation of point-of-care assays for glycated haemoglobin
Authors:St John Andrew  Davis Timothy M E  Goodall Ian  Townsend Mary Anne  Price Christopher P
Affiliation:ARC Consulting, Perth, W Australia 6050, Australia. stjohn@starwon.com.au
Abstract:
BACKGROUND: Various devices are now available to measure glycated haemoglobin (HbA1c) outside of the laboratory. The aim of this study was to assess the performance of these point-of-care instruments in the hands of non-laboratory trained personnel. METHODS: Two nursing staff tested samples from patients attending a diabetes research clinic using the following point-of-care devices for HbA1c-Metrica A1C Now, Bayer DCA 2000, Cholestech GDX and Axis-Shield Nycocard HbA1c. In addition they performed regular analysis of quality control samples. The effects on analytical performance of multiple operators as well as laboratory-trained staff, were also assessed. All measurements were compared to a boronate-affinity HPLC method in the central laboratory. RESULTS: The mean HbA1c difference of the point-of-care devices compared to the laboratory reference method ranged from -0.31% to +0.39%. Only the DCA device had a between batch imprecision of less than 5%. The analytical performance obtained by laboratory staff was similar to nursing staff for 3 devices and better for the Nycocard device. CONCLUSIONS: On the basis of the results obtained by nursing staff, only the DCA of the devices tested, can be recommended for measurement of HbA1c outside of the laboratory.
Keywords: