Dose and image-quality evaluation in synchrotron radiation mammography
Authors:E. Burattini  M. Gambaccini  P. L. Indovina  M. Marziani  S. Simeoni  A. Taibi
Affiliation:(1) Radiology Department of Verona University, I-37100 Verona, Italy;(2) INFN Laboratori Nazionali di Frascati, I-00044 Frascati, Rome, Italy;(3) Physics Department of Ferrara University, I-44100 Ferrara, Italy;(4) INFN Sez. of Ferrara, I-44100 Ferrara, Italy;(5) Physics Department of Naples University, I-80125 Naples, Italy;(6) INFN Sez. of Naples, I-80125 Naples, Italy
Abstract:
Monochromatic X-ray beam produced by synchrotron radiation may be considered an ideal probe in some fields of diagnostic radiology. In this paper the potential of monochromatic synchrotron radiation X-ray beam in mammography is analyzed. The image quality of four different phantom radiographs, obtained using two different energies equal to 17 and 18 keV, respectively, and a mammographic film/screen system, is assessed. The doses have been determined for both energies by using thermoluminescent dosimeters and a 5-cm thick phantom having a composition of 50% water and 50% fat. The results have been compared with those obtained in the same manner using a mammographic unit equipped with a molybdenum anode, molybdenum filter tube and antiscanner grid. The radiographs obtained with synchrotron radiation show higher contrast and less blur than those obtained with a conventional mammographic system. The average breast doses, 1.55 and 0.80 mGy at 17 and 18 keV, respectively, are comparable with the dose (1.51 mGy) of the conventional technique.Correspondence to: M. Gambaccini
Keywords:Mammography  Synchrotron radiation  Image quality  Phantom  Dosimetry
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号