Dose and image-quality evaluation in synchrotron radiation mammography |
| |
Authors: | E. Burattini M. Gambaccini P. L. Indovina M. Marziani S. Simeoni A. Taibi |
| |
Affiliation: | (1) Radiology Department of Verona University, I-37100 Verona, Italy;(2) INFN Laboratori Nazionali di Frascati, I-00044 Frascati, Rome, Italy;(3) Physics Department of Ferrara University, I-44100 Ferrara, Italy;(4) INFN Sez. of Ferrara, I-44100 Ferrara, Italy;(5) Physics Department of Naples University, I-80125 Naples, Italy;(6) INFN Sez. of Naples, I-80125 Naples, Italy |
| |
Abstract: | Monochromatic X-ray beam produced by synchrotron radiation may be considered an ideal probe in some fields of diagnostic radiology. In this paper the potential of monochromatic synchrotron radiation X-ray beam in mammography is analyzed. The image quality of four different phantom radiographs, obtained using two different energies equal to 17 and 18 keV, respectively, and a mammographic film/screen system, is assessed. The doses have been determined for both energies by using thermoluminescent dosimeters and a 5-cm thick phantom having a composition of 50% water and 50% fat. The results have been compared with those obtained in the same manner using a mammographic unit equipped with a molybdenum anode, molybdenum filter tube and antiscanner grid. The radiographs obtained with synchrotron radiation show higher contrast and less blur than those obtained with a conventional mammographic system. The average breast doses, 1.55 and 0.80 mGy at 17 and 18 keV, respectively, are comparable with the dose (1.51 mGy) of the conventional technique.Correspondence to: M. Gambaccini |
| |
Keywords: | Mammography Synchrotron radiation Image quality Phantom Dosimetry |