首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical evaluation of esophageal Doppler cardiac output measurement during general anesthesia
Authors:Mitsuo Ueda  Sho Yokota  Fusazo Nakata  Shigeo Kaseno  Norihiko Sakuraya  Osamu Kemmotsu
Affiliation:(1) Department of Anesthesiology, Hokkaido University School of Medicine, Sapporo, Japan;(2) Department of Anesthesiology, Hokkaido University School of Medicine, N-15, W-7, Kita-ku, Sapporo, 060, Japan
Abstract:
We evaluate the accuracy of cardiac output measurement with esophageal Doppler ultrasonography (ECO). A total of 71 simultaneous measurement of esophageal Doppler and thermodilution cardiac output were compared in 16 patients undergoing general anesthesia in the supine position. ECO was determined easily with minimal experience, and significantly correlated with thermodilution cardiac output (TDCO) measurement (P < 0.001). The regression equation obtained was Y = 0.983X + 0.019, and the correlation coefficient was 0.935. Furthermore, ECO was more reproducible than TDCO. However, ECO is not able to assess CO accurately in either lateral or prone position and after cardiopulmonary bypass in open heart surgery. Our results suggest that the esophageal Doppler technique allows a noninvasive and continuous cardiac output monitoring in patients during general anesthesia, and that it is more useful in patients for whom invasive monitoring is considered inappropriate. However, further improvement in this technique will be necessary for its routine use in clinical anesthesia.(Ueda M, Yokota S, Nakata F et al.: Clinical evaluation of esophageal doppler cardiac output measurement during general anesthesia. J Anesth 3: 178–182, 1989)
Keywords:Cardiac output  Esophageal Doppler ultrasonography  Thermodilution
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号