Influence of Body Position on Defibrillation Thresholds of Nonthoracotomy Implantable Defibrillators:
Authors:PATRICK SCHAUERTE,M.D.,,BJÖ  RN DIEM,M.D.,,KATHRIN ZIEGERT,M.D.,,REAS FRANKE,M.D.,,PETER HANRATH,M.D.,,CHRISTOPH STELLBRINK,M.D.
Affiliation:Department of Cardiology, University Hospital RWTH Aachen, Aachen, Germany
Abstract:
DFT of Nonthoracotomy Defibrillators. Introduction : Defibrillation thresholds (DFTs) usually are determined with the patient in the supine position. However, patients may be in the upright position when a shock is delivered during follow-up, which may explain some first shock failures observed clinically. This study investigated whether body posture affects defibrillation energy requirements of nonthoracotomy implantable cardioverter defibrillators with biphasic shocks.
Methods and Results : Using a step up-down protocol, DFTs were compared intraindividually in 52 patients ("active-can" systems in 41 patients, two-lead systems in II patients) for the supine and upright positions as achieved by a tilt table. The mean DFT was 7.3 ± 4.2 J in the supine versus 9.2 ± 4.8 J in the upright position (P = 0.002). Repeated comparison in reversed order 3 months after implantation in 22 patients revealed thresholds of 6.2 ± 2.5 J (supine) versus 8.4 ± 3.7 J (upright; P < 0.03) 1 week and 4.4 ± 2.4 J (supine) versus 6.2 ± 4.1 J (upright; P < 0.04) 3 months after implantation. DFTs decreased significantly for both body positions from 1 week to 3 months after implantation (P < 0.04).
Conclusion :(I) DFTs for biphasic shocks delivered by nonthoracotomy defibrillators are higher in the upright compared to the supine body position. (2) Differences remain significant 3 months after implantation. For both body positions, DFT decreases significantly from 1 week to 3 months after implantation. These findings have important implications for programming first shock energy to lower than maximal values or for development of devices with lower maximal stored energy.
Keywords:defibrillation threshold    implantable defibrillator    body position