4种国产和3种进口第4代HIV抗原抗体检测试剂的质量评价
引用本文:李秀华,许四宏,宋爱京,聂建辉,王佑春.4种国产和3种进口第4代HIV抗原抗体检测试剂的质量评价[J].中华微生物学和免疫学杂志,2011,31(9).
作者姓名:李秀华  许四宏  宋爱京  聂建辉  王佑春
作者单位:100050北京,中国食品药品检定研究院,卫生部生物技术产品检定方法及其标准化重点实验室
摘    要:目的 对4种国产和3种进口第4代HIV诊断试剂的质量进行评价.方法 利用HIV抗体阴性样品库和核酸阳性样品库、BBI阳转血清盘样品等,对4种国产和3种进口第4代试剂的敏感性和特异性、检测HIV-1早期感染的能力进行分析.结果 7种第4代试剂的敏感性均为100% (95% CI:99.86%~100%),且1份HIV-1感染窗口期样品均检测为阳性,其中1种进口试剂“8+”值较大(1.0892),其余6种试剂的“δ+”值均比较小(0.0836 ~0.3003).对阴性样品,7种试剂均存在不同程度的假阳性(特异性为97.80% ~ 99.60%,“δ-”值为-1.3803 -0.4778).对BBI阳转血清盘样品,国产试剂的阳转血清相对敏感性系数为-0.500~0,2种进口试剂则为-0.600和-0.700.结论 7种试剂均具有较高的敏感性和特异性,第4代HIV试剂用于血液筛查可发现HIV感染窗口期样本,对减少HIV传播的风险有一定的意义.但进口第4代试剂检测HIV-1早期感染的能力强于国产第4代试剂.

关 键 词:第4代HIV诊断试剂  HIV抗体阴性样品库  HIV核酸阳性样品库  BBI阳转血清  质量评价

Evaluation of four domestic and three imported fourth-generation HIV diagnostic reagents
LI Xiu-hua,XU Si-hong,SONG Ai-jing,NIE Jian-hui,WANG You-chun..Evaluation of four domestic and three imported fourth-generation HIV diagnostic reagents[J].Chinese Journal of Microbiology and Immunology,2011,31(9).
Authors:LI Xiu-hua  XU Si-hong  SONG Ai-jing  NIE Jian-hui  WANG You-chun
Abstract:Objective To evaluate the quality of four domestic and three imported fourth-generation HIV diagnostic reagents.Methods The specificity and sensitivity of these assays were analyzed when testing HIV negative samples and HIV-1 RNA positive samples.The relative seroconversion sensitivity index was analyzed when testing BBI seroconversion panels.Results The sensitivity of seven 4th-generation assays were 100% (95% CI:99.86%-100% ),and one sample at the window period of HIV-1 infection were detected as positive.Of the seven assays,one imported assay exhibited the relative large " δ +" value (1.0892),and the small "δ+" value were found on the remaining six assays (0.0836-0.3003 ).For the samples negative for HIV antibody,varying degrees of false positives were observed on the seven assays ( specificity:97.80% -99.60%,"δ-" value:-1.3803 to -0.4778).When testing the BBI seroconversion panels,the relative seroconversion sensitivity index of domestic assays were -0.500-0,however,which of imported assays were -0.600 and -0.700.Conclusion The seven reagents exhibited high sensitivity and specificity.The 4th generation HIV assays can be used as blood screening reagents to find the samples at window period of HIV-1 infection,thus indicating the certain meaning in reducing the transmission risk of HIV-1 for fourth-generation HIV diagnostic reagents.However,the better efficiency to detect HIV-1 early infection was observed on the imported assays than on the domestic assays.
Keywords:Fourth-generation HIV diagnostic reagents  HIV bank of negative samples  HIV bank of samples positive for nucleic acid  BBI seroconversion panel  Evaluation of quanlity