首页 | 本学科首页   官方微博 | 高级检索  
     

常用医疗诊断X射线机散漏射线剂量测量结果及讨论
引用本文:彭李青,朱志贤,何韦川,陈峰. 常用医疗诊断X射线机散漏射线剂量测量结果及讨论[J]. 中国辐射卫生, 2004, 13(3): 172-174. DOI: 10.13491/j.cnki.issn.1004-714x.2004.03.006
作者姓名:彭李青  朱志贤  何韦川  陈峰
作者单位:1. 深圳市坑梓人民医院, 广东深圳 518122;2. 深圳市疾病预防控制中心
摘    要:
目的 探讨X射线散漏射线对机房面积、设计参数的影响。方法 对8台常用医疗诊断X射线机的散漏射线的分布进行了实地测量。结果 散漏射线的减弱规律约为指数率,主要与X射线机本身性能有关。结论 在防护设计中应综合考虑X射线机性能,土地的商品价值,潜在的受照人群分布等多种因素,以求满足最优化要求。

关 键 词:医用X射线机  散漏射线  机房面积  
文章编号:1004-714X(2004)03-0172-03
收稿时间:2004-02-12
修稿时间:2004-02-12

Results and Discussions of the Dose Measurements of Scattering and Leaking Rays from Commonly Used Medical Diagnosis X-ray Machines
PENG Li-qing,ZHU Zhi-xian,HE Wei-chuan,et al. Results and Discussions of the Dose Measurements of Scattering and Leaking Rays from Commonly Used Medical Diagnosis X-ray Machines[J]. Chinese Journal of Radiological Health, 2004, 13(3): 172-174. DOI: 10.13491/j.cnki.issn.1004-714x.2004.03.006
Authors:PENG Li-qing  ZHU Zhi-xian  HE Wei-chuan  et al
Affiliation:Shenzhen Kengzi People's hospital, Shenzhen 518122, China
Abstract:
Objective To explore the influences of the scattering and leaking rays from X-ray machines on the area of the room of the machines installed and the design parameter. Methods Measurements of the distributions of scattering and leaking rays from eight types of medical diagnostic X-ray machines were made. Results The reduction of intensities of scattering and leaking rays obey approximately the index law and mainly depend on the performances of X-rays machine used. Conclusion The design of machine room should take such factors as the performance of the X-ray machine, the commodity value of the land and the distribution of people potentially affected, into potentially affected, into consideration in order to realize optimization of radiation protection.
Keywords:Medical X-ray Machine  Scattering and Leaking Rays  Area of Machine Room.
本文献已被 CNKI 维普 万方数据 等数据库收录!
点击此处可从《中国辐射卫生》浏览原始摘要信息
点击此处可从《中国辐射卫生》下载全文