用二维电离室阵列对螺旋断层治疗的调强计划进行剂量验证
引用本文:徐寿平,解传滨,鞠忠建,戴相昆,郭妍妍,巩汉顺,王连元. 用二维电离室阵列对螺旋断层治疗的调强计划进行剂量验证[J]. 中华放射肿瘤学杂志, 2000, 18(1): 233-236. DOI: 10.3760/cma.j.issn.1004-4221.2009.03.233
作者姓名:徐寿平  解传滨  鞠忠建  戴相昆  郭妍妍  巩汉顺  王连元
作者单位:解放军总医院放疗科,100853;天津医科大学附属二院放疗科;
摘    要:Objective To investigate the feasibility of dose verification of intensity modulated (IM) planning of helical tomotherapy (HT) using two-dimensional ion chamber array (2DICA),and develop an efficient way to validate the dose delivered under the parameters mirroring those during the treatment. Meth-ods A 2DICA,I'mRT MatriXX and MULTICube equivalent solid water phantom from IBA company were used to verify the dose distribution of 10 IM planning. The combined phantom was set up to measure the dose distributions on coronal and sagittal surface. The precise setup of phantom was guided by HTMVCT images. After the irradiation, the measured dose distributions on the coronal and sngittal plane were compared with those calculated by the IM planning system for verification. The results were evaluated and the feasibility of the different measuring methods was studied. Results The dose distribution measured by the MatriXX 2DICA was well consistent with that calculated by the treatment planning system. The errors between the measured dose and predicted dose in the selected points were within ±3%. In the comparison of the pixel-segmented ionization chamber versus treatment planning system using the 3 mm/3% γ criteria, the passing ratio of pixels with γ parameter ≤1 was 97.76% and 96.83%, respectively. Conclusions MatriXX is a-ble to measure the absolute and relative dose distributions simultaneously,which can be used for dose verifi-cation of IM planning.

关 键 词:螺旋断层治疗   二维电离室阵列   等效固体水模   剂量验证   

Two-dimensional ion chamber array in dose verification for intensity modulated planning of helical tomotherapy
XU Shou-ping,XIE Chuan-bin,JU Zhong-jian,DAI Xiang-kun,GUO Yan-yan,GONG Han-shun,WANG Lian-yuan. Two-dimensional ion chamber array in dose verification for intensity modulated planning of helical tomotherapy[J]. Chinese Journal of Radiation Oncology, 2000, 18(1): 233-236. DOI: 10.3760/cma.j.issn.1004-4221.2009.03.233
Authors:XU Shou-ping  XIE Chuan-bin  JU Zhong-jian  DAI Xiang-kun  GUO Yan-yan  GONG Han-shun  WANG Lian-yuan
Abstract:
Keywords:Helical tomotherapyTwo-dimensional ion chamber arrayEquivalent solid water phantomDose verification