首页 | 本学科首页   官方微博 | 高级检索  
     


Enzymatic determination of serum ethanol with membrane-bound dehydrogenase
Authors:H Takei  K Nakashima  O Adachi  E Shinagawa  M Ameyama
Abstract:
In this enzymatic method for detecting ethanol in blood by use of membrane-bound microbial alcohol dehydrogenase (no EC no. assigned), the enzyme catalyzes the reaction irreversibly and the rate of oxidation can be monitored by spectrophotometry of the reduction of the indicator dye. No pyridine nucleotides such as NAD+ or NADP+ are used. The calibration curve is linear in the range of 0.1 to 4.0 g of ethanol per liter. Assays of 45 samples of serum having ethanol values ranging from 0.4 to 3.2 g/L by the described technique and a gas-chromatographic method gave respective means of 1.734 and 1.732 g/L (r = 0.954).
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号