Double-phase 99mTc-sestamibi scintimammography and trans-scan in diagnosing breast cancer. |
| |
Authors: | M Melloul A Paz G Ohana O Laver D Michalevich R Koren Y Wolloch R Gal |
| |
Affiliation: | Department of Nuclear Medicine, Hasharon Hospital, Rabin Medical Center, Petah Tikva, Israel. |
| |
Abstract: | The goal of our study was to assess the value of both scintimammography with 99mTc-sestamibi (SMM) and trans-scan (T-scan) in detecting breast cancer. METHODS: A total of 121 women were evaluated by palpation, mammography, SMM and T-scan. SMM was performed in the prone, breast dependent position. Immediate and delayed views (double-phase) were obtained. T-scan is a new breast imaging method that maps noninvasively the distribution of tissue electrical impedance and capacitance. RESULTS: SMM had 88.9% sensitivity, 88.4% specificity and 88.4% accuracy in detecting breast cancer. SMM had 100% sensitivity in detecting breast tumors >1 cm and only 66% sensitivity in detecting tumors <1 cm. T-scan had 72.2% sensitivity and 67% specificity in detecting breast cancer. It detected one more breast cancer than SMM, at the expense of 27 additional false-positive results. CONCLUSION: Double-phase SMM was sensitive and specific in detecting breast cancer. This method may reduce the rate of negative breast biopsies in tumors >1 cm. T-scan was only moderately accurate in detecting breast cancer. Its addition to SMM did not improve significantly the rate of breast cancer detection. However, because of its complete noninvasiveness, large-scale applicability and low cost, T-scan deserves further refining. |
| |
Keywords: | |
|
|