首页 | 本学科首页   官方微博 | 高级检索  
     


Normal and stenotic renal arteries: experimental balloon-expandable intraluminal stenting
Authors:Palmaz, JC   Kopp, DT   Hayashi, H   Schatz, RA   Hunter, G   Tio, FO   Garcia, O   Alvarado, R   Rees, C   Thomas, SC
Abstract:Elastic recoil of the vessel wall is a common cause of failure of percutaneous transluminal angioplasty in renal arteries. To oppose such recoil, balloon-expandable metal stents were implanted in artificially stenotic renal arteries in pigs and normal renal arteries in dogs and pigs. The stents were then examined angiographically and histologically at regular intervals. All stents were completely covered with endothelialized neointima in 3 weeks. There was no difference in intimal thickness between the stenotic and nonstenotic renal arteries. A large stent diameter and a large open or nonmetal surface may cause less intimal hyperplasia, but nonturbulent, fast arterial flow is probably the most important factor in ensuring long-term patency of the vessel.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号