首页 | 本学科首页   官方微博 | 高级检索  
     


Modified positioning of a smartphone based single-lead electrocardiogram device improves detection of atrial flutter
Authors:Kevin Rajakariar  Anoop N. Koshy  Jithin K. Sajeev  Sachin Nair  Louise Roberts  Andrew W. Teh
Affiliation:1. Department of Cardiology, Monash University, Eastern Health Clinical School, Victoria, Australia;2. Department of Cardiology, The University of Melbourne, Austin Hospital Clinical School, Melbourne, Australia;3. Monash University, Eastern Health Clinical School, Victoria, Australia
Abstract:

Introduction

The AliveCor Kardia Mobile (AKM) is a handheld, smartphone based cardiac rhythm monitor that records a lead-I electrocardiogram (ECG). Despite being efficacious for detection of atrial fibrillation (AF), it is unclear whether atrial flutter (AFL) may be misdiagnosed as sinus rhythm due to regular R-R intervals. We hypothesised that generating lead-II tracings through repositioning of the AKM may improve visualisation of flutter waves and clinician diagnosis of AFL compared to traditional lead-I tracings.

Materials and methods

A prospective, multi-centre, validation study was conducted comparing standard lead-I AKM positioning with lead-II in AFL. A mixed cohort of lead I tracings from patients in AF and sinus rhythm were also included. Two independent electrophysiologists (EP) analysed all ECGs blinded to the automated device diagnosis.

Results

Fifty patients were recruited, 11 in atrial flutter, 14 in atrial fibrillation, and 25 in sinus rhythm. Lead-I AFL sensitivity was 27.3% for both EP's which individually improved to 72.7% and 54.6% in lead-II. AKM appropriately diagnosed lead-I AFL as unclassified in 18.2% of cases, compared to 54.5% in lead-II. Overall clinician agreement (AF, SR and AFL) was modest utilising AFL lead-I (EP1: κ?=?0.71, EP2: κ?=?0.73, p?

Conclusion

Repositioning of the AKM device improves clinician diagnosis of atrial flutter. A lead-II tracing may be considered in high-risk patients to improve detection of atrial flutter.
Keywords:AliveCor  Kardia Mobile  Atrial flutter  Wireless  Smartphone
本文献已被 ScienceDirect 等数据库收录!
正在获取相似文献,请稍候...
设为首页 | 免责声明 | 关于勤云 | 加入收藏