Measurement of the dose equivalent of leakage radiation through an isocentric gantry used for neutron therapy
Authors:J L Horton  P M Stafford  P Kliauga
Affiliation:Department of Radiation Physics, University of Texas M.D. Anderson Hospital and Tumor Institute, Houston 77030.
Abstract:
The leakage radiation through the shielding on an isocentric gantry of a neutron therapy machine was measured with a Rossi-type proportional counter. The dose equivalent of the leakage radiation was determined at two positions: (1) in the plane of the patient and (2) in the plane of the target. The dose equivalent of the leakage radiation is approximately the same as the leakage of a high-energy x-ray linac.
Keywords: