两种高密度脂蛋白胆固醇测定试剂盒的性能评价
引用本文:熊晓芸,徐光,朱宝立. 两种高密度脂蛋白胆固醇测定试剂盒的性能评价[J]. 江苏预防医学, 2014, 25(6): 3-5
作者姓名:熊晓芸  徐光  朱宝立
作者单位:江苏省疾病预防控制中心,南京,210009
基金项目:卫生部卫生行业科研专项,江苏省医学领军人才与创新团队项目
摘    要:目的比较免疫抑制直接法(AB法)与选择性抑制法(PPD法)两种试剂盒,评价其在测定高密度脂蛋白(HDL-C)上的性能。方法在全自动生化分析仪上,分别检测两种试剂的精密度、测定范围、抗干扰能力,以及临床测定标本结果的相关性。结果 AB法试剂和PPD法试剂的批内精密度均1%,批间精密度均2%,均有较好的精密度。AB法试剂测定范围为0.04~4.0mmol/L,PPD法试剂为0.06~2.60mmol/L。AB法试剂和PPD法试剂在干扰物质Hb5.0g/L时,对结果没有明显影响,在干扰物质TG13.0mmol/L时,用AB法试剂测定HDL-C没有明显影响,PPD法试剂在干扰物质TG≥13.0mmol/L时,检测结果偏差15%。两种试剂测量临床样本具有良好的相关性,回归方程为Y=0.867 X+0.164(r=0.982,P0.05)。结论 AB法试剂抗干扰能力优于PPD法试剂;两种试剂测定方法具有良好的相关性,符合临床检测要求。

关 键 词:高密度脂蛋白  免疫抑制直接法  选择性抑制法

Performance evaluation of two kinds of determination kit for high density li-poprotein cholesterol
XION Xiao-yun,XU Guang,ZHU Bao-li. Performance evaluation of two kinds of determination kit for high density li-poprotein cholesterol[J]. Jiangsu Journal of Preventive Medicine, 2014, 25(6): 3-5
Authors:XION Xiao-yun  XU Guang  ZHU Bao-li
Affiliation:(J iangsu Province Center for Disease Control and Prevention, Nanj ing 210028, China)
Abstract:Objective To compare two kits used respectively on the immune inhibitor direct method and selective inhibitor direct method so as to evaluate the performance on measuring high density lipoprotein cholesterol(HDL-C).Methods On fullautomatic biochemical analyzer,the A kit measuring method is immune inhibitor direct method and the B kit measuring method is selective inhibitor direct method.The precision,measuring range,anti-interference capability and the correlation of the results of clinical specimens were measured by two kits respectively.Results The within run precision and between run precisionof A kit and B kit are both 〈1% and 〈2%,respectively.The two kits have both preferable precision.The measuring range of A kit and B kit are 0.04~4.0mmol/L and 0.06~2.60mmol/L,respectively.The analyzed results by two kits have not obvious impact when interfering substance Hb〈5.0g/L.The HDL-C results by A kit have not obvious impact when interfering substance TG〈13.0 mmol/L.The bias of analyzed results by B kit reached 15% when interfering substance TG≥13.0mmol/L.The correlation of the results of clinical specimens of two kits was preferable(Regression equation Y=0.867X+0.164;r=0.982).Conclusion The anti-interference capability of A kit was better than that of B kit.The measuring methods by two kits were preferable comparability.They all meet the needs of clinical detection of specimens.
Keywords:high density lipoprotein cholesterol  immune inhibitor direct method  selective inhibitor direct method