首页 | 本学科首页   官方微博 | 高级检索  
     


Multicenter analytical evaluation of an automated immunoassay for total plasma homocysteine
Authors:Wu A H  Holtman V  Apple F S  Ricchiuti V  DiBello P M  Jacobsen D
Affiliation:Department of Pathology and Laboratory Medicine, Hartford Hospital, Connecticut 06102, USA. awu@harthosp.org
Abstract:
A fully automated immunoassay for total plasma homocysteine assay was evaluated at four centers. To measure total homocysteine, oxidized forms of homocysteine in serum and plasma were reduced by dithiothreitol and assayed by a competitive fluorescence polarization technique. The assay had within-run precision from 0.9 to 3.0% and total precision from 2.8 to 4.1% for control materials with homocysteine concentrations of approximately 7, 12.5, and 25 micromol/L, a sensitivity of 0.35 micromol/L, good parallelism upon dilution, and analytical recovery ranging from 97.4 to 103.8%. The immunoassay correlated with four different HPLC assays for homocysteine, yielding a slope of 0.98, an intercept of -0.19 micromol/L, and a correlation coefficient of 0.966 for 440 paired samples. The reference range, determined with plasma samples from 609 males and 600 females, yielded a mean of 9.17+/-2.86 micromol/L, with a central 95% range of 4.78-15.43 micromol/L. The immunoassay is a suitable alternative to HPLC and may be useful in screening persons with high risk of coronary artery disease.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号