首页 | 本学科首页   官方微博 | 高级检索  
     


Megahertz streak-mode Fourier domain optical coherence tomography
Authors:Wang Rui  Yun Julie X  Yuan Xiaocong  Goodwin Richard  Markwald Roger R  Gao Bruce Z
Affiliation:Clemson University, Department of Bioengineering, COMSET, Clemson, South Carolina 29634, USA.
Abstract:
Here we present an ultrahigh-speed Fourier-domain optical coherence tomography (OCT) that records the OCT spectrum in streak mode with a high-speed area scan camera, which allows higher OCT imaging speed than can be achieved with a line-scan camera. Unlike parallel OCT techniques that also use area scan cameras, the conventional single-mode fiber-based point-scanning mechanism is retained to provide a confocal gate that rejects multiply scattered photons from the sample. When using a 1000 Hz resonant scanner as the streak scanner, 1,016,000 A-scans have been obtained in 1 s. This method's effectiveness has been demonstrated by recording in vivo OCT-image sequences of embryonic chick hearts at 1000 frames/s. In addition, 2-megahertz OCT data have been obtained with another high speed camera.
Keywords:
本文献已被 PubMed 等数据库收录!