Impedance of Pacemaker Leads: Correlation of Different Methods |
| |
Authors: | JOACHIM B. SIEGMUND JOHN H. WILSON STEPHEN E. LATTNER KIMBERLY GRANNEMAN ROBERT JOHNSON |
| |
Affiliation: | Department of Cardiology, Good Samaritan Hospital, Cincinnati, Ohio |
| |
Abstract: | This study compares three different pacing system analyzers, which measure impedance at different points during an impulse, with measurements telemetered from implanted pacemakers from four different manufacturers. Measurements were obtained at the time of implantation in a group of 103 patients. The measurements obtained by these different methods differ significantly from each other; the later during the impulse the impedance is measured, the higher in general its value. Interpretation of impedance values should take into account the technique of measurement. |
| |
Keywords: | lead impedance pacing system analyzer polarization effect |
|
|