Impedance of Pacemaker Leads: Correlation of Different Methods
Authors:JOACHIM B. SIEGMUND  JOHN H. WILSON  STEPHEN E. LATTNER  KIMBERLY GRANNEMAN  ROBERT JOHNSON
Affiliation:Department of Cardiology, Good Samaritan Hospital, Cincinnati, Ohio
Abstract:
This study compares three different pacing system analyzers, which measure impedance at different points during an impulse, with measurements telemetered from implanted pacemakers from four different manufacturers. Measurements were obtained at the time of implantation in a group of 103 patients. The measurements obtained by these different methods differ significantly from each other; the later during the impulse the impedance is measured, the higher in general its value. Interpretation of impedance values should take into account the technique of measurement.
Keywords:lead impedance    pacing system analyzer    polarization effect