探讨视觉质量分析系统(OQAS)测量圆锥角膜眼与正常角膜眼之间视觉质量的差异。方法:横断面研究。选取正常角膜屈光不正患者32例(32眼)作为对照组,圆锥角膜患者27例(27眼)作为圆锥角膜组。采用OQAS测量每组在3、4、5 mm的人工瞳孔直径下的视觉质量参数调制传递函数截止频率(MTFcutoff)、Strehl比、不同对比度下的OQAS值(OV-100、OV-20、OV-9),组间差异采用独立样本t检验或Mann-Whitney U检验,组内差异采用重复测量资料方差分析及Bonferroni校正。结果:圆锥角膜组在人工瞳孔分别为3、4、5 mm时的MTFcutoff、Strehl比、OV-100、OV-20、OV-9均低于对照组(3 mm:t=6.658、6.695、6.660、6.493、6.778;4 mm:t=7.106、6.567、7.109、6.802、6.788; 5 mm:t=6.148、7.254、6.158、6.446、7.210;P均<0.001)。对照组及圆锥角膜组内MTFcutoff、Strehl比、 OV-100、OV-20、OV-9 在不同人工瞳孔大小的组内比较差异有统计学意义(对照组:F=42.205、 28.335、42.559、32.621、30.372;圆锥角膜组:F=15.732、11.533、15.625、13.870、10.495;均 P<0.001)。结论:双通道OQAS测量结果表明圆锥角膜的视觉质量较正常眼明显下降,圆锥角膜及正常眼的视觉质量均随着瞳孔增大而下降。

关 键 词:圆锥角膜  视觉质量  瞳孔  双通道  
收稿时间:2019-01-14

Objective Assessment of Optical Quality in Mild Keratoconus with a Double-Pass System
Jia Zhang,Yong Chen,Wenwen Zhao,Chenwei Wu,Chenchen Xu,Ling Wang,Shihao Chen. Objective Assessment of Optical Quality in Mild Keratoconus with a Double-Pass System[J]. Chinese Journal of Optometry Ophthalmology and Visual Science, 2019, 21(12): 900-905. DOI: 10.3760/cma.j.issn.1674-845X.2019.12.004
Authors:Jia Zhang  Yong Chen  Wenwen Zhao  Chenwei Wu  Chenchen Xu  Ling Wang  Shihao Chen
Affiliation:Eye Hospital, Wenzhou Medical University, Wenzhou 325027, China
Abstract:
To analyze the difference in optical quality between a normal cornea and keratoconus using an optical quality analysis system (OQAS). Methods: This cross-sectional study included 32 normal eyes and 27 keratoconus eyes. Optical qualities were measured by OQAS. Modulation transfer function cutoff frequency (MTFcutoff), Strehl ratio and OQAS values (OV-100, OV-20, OV-9) in artificial pupil diameters of 3 mm, 4 mm and 5 mm were analyzed by independent t test or Mann-Whitney U test, the difference among groups was analyzed by ANOVA for repeated measurement and Bonferroni post hoc t test. Results: MTFcutoff, Strehl ratio and OQAS values of the keratoconus group in artificial pupil diameters of 3 mm, 4 mm and 5 mm were all lower than those of the control group (3 mm: t=6.658, 6.695, 6.660, 6.493, 6.778; 4 mm: t=7.106, 6.567, 7.109, 6.802, 6.788; 5 mm: t=6.148, 7.254, 6.158, 6.446, 7.210; all P<0.001). Both in the control group and in the keratoconus group, the MTFcutoff, Strehl ratio, OV-100, OV-20 and OV-9 varied from artificial pupil diameters (control group: F=42.205, 28.335, 42.559, 32.621, 30.372; keratoconus group: F=15.732, 11.533, 15.625, 13.870, 10.495; all P<0.001). Conclusions: The optical qualities of keratoconus measured by OQAS are worse than normal eyes. Both in keratoconus and normal eyes, the optical qualities decline as pupil size increases.
Keywords:keratoconus   optical quality  pupil  double-pass  