Comparison of quantitative perfusion imaging using arterial spin labeling at 1.5 and 4.0 Tesla. |
| |
Authors: | Jiongjiong Wang David C Alsop Lin Li John Listerud Julio B Gonzalez-At Mitchell D Schnall John A Detre |
| |
Affiliation: | Department of Radiology, University of Pennsylvania, Philadelphia 19104, USA. |
| |
Abstract: | High-field arterial spin labeling (ASL) perfusion MRI is appealing because it provides not only increased signal-to-noise ratio (SNR), but also advantages in terms of labeling due to the increased relaxation time T(1) of labeled blood. In the present study, we provide a theoretical framework for the dependence of the ASL signal on the static field strength, followed by experimental validation in which a multislice pulsed ASL (PASL) technique was carried out at 4T and compared with PASL and continuous ASL (CASL) techniques at 1.5T, both in the resting state and during motor activation. The resting-state data showed an SNR ratio of 2.3:1.4:1 in the gray matter and a contrast-to-noise ratio (CNR) of 2.7:1.1:1 between the gray and white matter for the difference perfusion images acquired using 4T PASL, 1.5T CASL, and 1.5T PASL, respectively. However, the functional data acquired using 4T PASL did not show significantly improved sensitivity to motor cortex activation compared with the 1.5T functional data, with reduced fractional perfusion signal change and increased intersubject variability. Possible reasons for these experimental results, including susceptibility effects and physiological noise, are discussed. |
| |
Keywords: | arterial spin labeling (ASL) cerebral blood flow (CBF) magnetic field strength multislice perfusion fMRI susceptibility effect |
|
|