BC-3000血细胞分析仪计数血小板及其参数与XT-2000i血细胞分析仪差异分析 |
| |
引用本文: | 华丽姿,郭秀荣.BC-3000血细胞分析仪计数血小板及其参数与XT-2000i血细胞分析仪差异分析[J].中国现代医生,2013,51(1):72-73,76. |
| |
作者姓名: | 华丽姿 郭秀荣 |
| |
作者单位: | 华丽姿 (辽宁省朝阳市建平县医院检验科,辽宁建平,122499);郭秀荣 (辽宁省朝阳市建平县医院检验科,辽宁建平,122499); |
| |
摘 要: | 目的探讨迈瑞BC-3000三分类血细胞分析仪与SysmexXT-2000i五分类血细胞分析仪计数血小板及其参数MPV、PDW、PCT差异性。方法收集血小板计数正常、增高和减低的患者标本共433例.分别用迈瑞BC-3000和SysmexXT-2000i两部血细胞计数仪测定PLT及MPV、PDW、PCT。结果迈瑞BC-3000血细胞分析仪计数正常组PLT及MPV、PDW、PCT与SysmexXT-2000i血细胞分析仪无显著差异(P〉0.05);计数血小板增多组的PLT、MPV、PDW、PCT四项指标有显著差异(P〈0.05);计数血小板减少组的PLT、MPV和PDW有显著差异(PLT,P〈0.05;MPV和PDW,P〈0.01),计数PCT两仪器无显著差异(P〉0.05)。结论迈瑞BC-3000血细胞分析仪计数血小板异常者,应用SysmexXT-2000i血细胞分析仪对其结果进行复核及校正。
|
关 键 词: | 血细胞分析仪 血小板计数 差异 |
Analysis of difference between BC-3000 blood cell analyzer and XT-2000i blood cell analyzer in calculating platelet and its parameters |
| |
Authors: | HUA Lizi GUO Xiurong |
| |
Institution: | Department of Clinical Laboratory, Jianping County Hospital in Liaoning Province, Jianping 122499, China |
| |
Abstract: | Objective To investigate the difference between the Mindray BC-3000 three-category blood cell analyzer and the Sysmex XT-2000i five-category blood cell analyzer in calculating the platelet (PLT) and its parameters mean platelet volume (MPV), platelet distribution width (PDW) and plateletcrit (PCT). Methods A total of 433 cases of nor- mal, increased and decreased platelet count samples were collected, of which the PLT, MPV, PDW and PCT were measured using the Mindray BC-3000 three-category blood cell analyzer and the Sysmex XT-2000i five-category blood cell analyzer respectively. Results In the normal platelet count group, the Mindray BC-3000 blood cell analyzer and the Sysmex XT-2000i blood cell analyzer were not significantly different in the PLT, MPV, PDW and PCT (P 〉 0.05); In the increased platelet count group, the two analyzers were significantly different in the PLT, MPV, PDW and PCT (P 〈 0.05); In the decreased platelet count group, the two analyzers were significantly different in the PLT, MPV and PDW (P 〈 0.05 for PLT; P 〈 0.01 for MP~ and PDW), but were not significantly different in the PCT (P 〉 0.05). Conclusion The Mindray BC-3000 blood cell analyzer is used to calculate the platelet of patients with abnormal platelet count and the Sysmex XT-2000i blood cell analyzer is used to recheck and revise its results. |
| |
Keywords: | Blood cell analyzer Platelet count Difference |