Heart rate measurement based on a time-lapse image
Authors:Takano Chihiro  Ohta Yuji
Affiliation:Graduate School of Humanities and Sciences, Ochanomizu University, Tokyo, Japan. c_takano@a.toshima.ne.jp
Abstract:
Using a time-lapse image acquired from a CCD camera, we developed a non-contact and non-invasive device, which could measure both the respiratory and pulse rate simultaneously. The time-lapse image of a part of the subject's skin was consecutively captured, and the changes in the average image brightness of the region of interest (ROI) were measured for 30s. The brightness data were processed by a series of operations of interpolation as follows a first-order derivative, a low pass filter of 2 Hz, and a sixth-order auto-regressive (AR) spectral analysis. Fourteen sound and healthy female subjects (22-27 years of age) participated in the experiments. Each subject was told to keep a relaxed seating posture with no physical restriction. At the same time, heart rate was measured by a pulse oximeter and respiratory rate was measured by a thermistor placed at the external naris. Using AR spectral analysis, two clear peaks could be detected at approximately 0.3 and 1.2 Hz. The peaks were thought to correspond to the respiratory rate and the heart rate. Correlation coefficients of 0.90 and 0.93 were obtained for the measurement of heart rate and respiratory rate, respectively.