Properties of implanted electrodes for functional electrical stimulation |
| |
Authors: | Dejan Popovic Tessa Gordon Victor F. Rafuse Dr. Arthur Prochazka |
| |
Affiliation: | (1) Division of Neuroscience, University of Alberta, T6G 2S2 Edmonton, Alta, Canada |
| |
Abstract: | Implanted wire electrodes are increasingly being used for the functional electrical stimulation of muscles in partially paralysed patients, yet many of their basic characteristics are poorly understood. In this study we investigated the selectivity, recruitment characteristics and range of control of several types of electrode in triceps surae and plantaris muscles of anaesthetized cats. We found that nerve cuffs are more efficient and selective (i.e., cause less stimulus spread to surrounding muscles) than intramuscular electrodes. Bipolar intramuscular stimulation was more efficient and selective than monopolar stimulation, but only if the nerve entry point was between the electrodes. Monopolar electrodes are efficient and selective if located close to the nerve entry point, but their performance declines with distance from it. Nonetheless, for a variety of reasons monopolar stimulation provides the best compromise in many current applications. Short duration pulses offer the best efficiency (least charge per pulse to elicit force) but high peak currents, increasing the risk of electrode corrosion and tissue damage. Electrode size has little effect on recruitment and should therefore be maximised because this minimises current density. |
| |
Keywords: | Electrical stimulation electrodes |