Atrial Synchronous Ventricular Pacing with a Single Lead: Reliability of Atrial Sensing During Physical Activities, and Long-term Stability of Atrial Sensing
Authors:SVEIN FAERESTRAND,OLE-JØ  RGEN OHM
Affiliation:Medical Department A, University of Bergen, Bergen, Norway
Abstract:
A VDD pacing system with bipolar single-pass leads, were implanted in 36 consecutive patients (average age 72 ± 2years) with high degree atrioventricular block and normal sinus node function. At implant the atrial signal amplitude was 2.6 ± 0.2mV measured by a pacing system analyser (PSA), 1.8 ± 0.1mV measured peak-to-peak from the telemetered calibrated electrogram, and 1.3 ± 0.1mV measured from the sensing threshold. At one month follow-up the peak-to-peak amplitudes (mV) of the telemetered atrial electrograms were not significantly different measured continuously during resting supine with quiet breathing (1.4 ± 0.1), sitting (1.6 ± 0.2). standing (1.5 ± 0.1), arm swinging (1.4 ± 0.2), hyperventilation (1.3 ± 0.1), Vaisalva manoeuvre (1.4 ± 0.1), and treadmill exercise (1.9 ± 0.6). The telemetered atrial electrogram amplitude and the atrial sensing threshold varied between 1.2 ± 0.09mV and 1.8 ± 0.1mV, and between 0.95 ± 0.07mV and 1.3 ± 0.01mV, respectively at 0.5, 1, 3, 6 and 12 months follow-up, but the changes were statistically nonsignificant. The Event Summary showed sensing of 98% to 99% of the atrial events at the different follow-up periods.
Keywords:single-pass lead    atrial sensing    long-term stability