Abstract:
In this paper the clinical introduction of stereoscopic kV-imaging in combination with a 6 degrees-of-freedom (6 DOF) robotics system and breathing synchronized irradiation will be discussed in view of optimally reducing interfractional as well as intrafractional geometric uncertainties in conformal radiation therapy. Extracranial cases represent approximately 70% of the patient population on the NOVALIS treatment machine (BrainLAB A.G., Germany) at the AZ-VUB, which is largely due to the efficiency of the real-time positioning features of the kV-imaging system. The prostate case will be used as an example of those target volumes showing considerable changes in position from day-to-day, yet with negligible motion during the actual course of the treatment. As such it will be used to illustrate the on-line target localization using kV-imaging and 6 DOF patient adjustment with and without implanted radio-opaque markers prior to treatment. Small lung lesion will be used to illustrate the system's potential to synchronize the irradiation with breathing in coping with intrafractional organ motion.