两种机型的流式细胞仪测定CD4T细胞计数的比较研究
引用本文:彭虹,徐建青,段丹丽,袁霖,绳波,王露,阮玉华,韩丽凤,苏斌,徐臣,邢辉,邵一鸣.两种机型的流式细胞仪测定CD4T细胞计数的比较研究[J].中国艾滋病性病,2006,12(3):199-201.
作者姓名:彭虹  徐建青  段丹丽  袁霖  绳波  王露  阮玉华  韩丽凤  苏斌  徐臣  邢辉  邵一鸣
作者单位:1. 中国疾病预防控制中心,性病艾滋病预防控制中心,北京,100050
2. 安徽省疾病预防控制中心,合肥,230061
3. 阜阳市疾病预防控制中心,安徽,阜阳,236000
基金项目:中国CIPRA项目;国家科技攻关项目
摘    要:目的多种流式细胞仪机型应用于艾滋病防治工作中,比较不同机型间的免疫数据乃是当前迫切需要解决的问题。方法应用BDFACS Calibur和Beckman Coulter XL两种主流机型流式细胞仪,对阜阳市现场采集的183份标本正常人41份,艾滋病病毒(HIV)阳性感染者142份]进行平行检测,观察上述两种机型对检测同一标本CD4绝对计数的影响。结果BD和XL两种机型的绝对计数结果比较显示,正常人CDd绝对计数在两种机型间的差异范围:≤10%的占57%,而〉10%占43%;检测142份HIV阳性标本,CD4绝对计数差异≤10%的占84%;而差异〉10%的占16%。在正常人标本组两机型数据间相关系数r=0.97;HIV阳性标本组数据间相关系数r=0.98,回归方程X系数均接近1。统计学检验显示,两种机型检测数据无显著性差异(P〉0.05)。结论两种机型间数据高度相关且数值相近.所测定的HIV感染者CD4细晌绡对计数可以百用.

关 键 词:艾滋病病毒  CD4细胞绝对计数  流式细胞仪
文章编号:1672-5662(2006)03-0199-03
收稿时间:2005-09-26
修稿时间:2006-03-22

Comparison of CD4T cell counts measured by flow cytometry of BD FACS Calibur and Beckman Coulter XL
PENG Hong, XU Jian-qing, DUAN Dan-li,et al..Comparison of CD4T cell counts measured by flow cytometry of BD FACS Calibur and Beckman Coulter XL[J].Chinese JOurnal of Aids & STD,2006,12(3):199-201.
Authors:PENG Hong  XU Jian-qing  DUAN Dan-li  
Institution:Division of Research on Virology and Immunology, National Center for AIDS / STD Control and Prevention, China CDC , Bering 100050, China
Abstract:Objective Since flow cytometry has been widely used in the national effort to contain AIDS epidemic in China,it is important and urgent to understand and compare CD_4 T-cell counts determined by different types of this apparatus.Methods To gain information in this aspect,we used two types of flow cytometry,i.e.,BD FACS Calibur and Beckman Coulter Xl in parallel,to quantify CD_4T-cell counts in 183 blood samples from Anhui province,including 41 samples from seronegative individuals and 142 samples from HIV-infected subjects.Result Overall,the difference in 57% of samples from 41 seronegative individuals was<10% while in 43% of samples it was>10%.In contrast,119 samples out of 142 ones(84%)in HIV-infected individuals generated<10% difference by using these two types of flow cytometry,whereas only 16% of samples produced >10% discrepancy.The correlation coefficient was equal to 0.97 in the seronegative group and 0.98 in the HIV-infected group.Statistical analysis showed no significant difference between data generated by these two types of flow cytometry(P>0.05).Conclusion Data generated by these two types of flow cytometry are highly correlated and can be directly used at a range of CD_4T-cell counts between 50 and 500.
Keywords:HIV infection  CD4T-cell counts  Calibur