首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of Postoperative Complications Following Metal-on-Metal Total Hip Arthroplasty With Other Hip Bearings in Medicare Population
Authors:Lindsay T. Kleeman  Abiram Bala  Colin T. Penrose  Thorsten M. Seyler  Samuel S. Wellman  Michael P. Bolognesi
Affiliation:1. Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina;2. Department of Orthopaedic Surgery, Stanford University Medical Center, Palo Alto, California
Abstract:

Background

The use of metal-on-metal (MoM) hip bearings has declined in the recent years due to strong evidence of their high complication rates and early failure. Hip implants with highly cross-linked polyethylene liners and ceramic bearings have become the modern implants of choice. We sought to determine if MoM implants are associated with higher complication and revision rates when compared to other hip bearings in the Medicare population.

Methods

We retrospectively reviewed a Medicare database (2005-2011) for patients who underwent a primary total hip arthroplasty with a MoM, metal-on-polyethylene (MoP), ceramic-on-polyethylene (CoP), or ceramic-on-ceramic (CoC) implant (minimum 2 years of follow-up). Patient comorbidities and medical/surgical complication rates were analyzed at various time points postoperatively.

Results

We identified 288,118 patients, including 81,520 patients with a MoM implant, 162,881 with MoP, 33,819 with CoP, and 9898 with CoC implant. Surgical complication rates were higher for MoM implants including infection, osteolysis/polywear, mechanical complications, and need for hip irrigation and debridement. Overall revision rates were significantly higher for MoM implants (5.28%) compared to MoP (4.28%, odds ratio [OR] 1.26, P < .001) and CoP (3.52%, OR 1.55, P < .001) but only by one to two percent. MoM revision rates were similar to CoC implants (4.94%, OR 1.00, P = .096).

Conclusions

MoM implants were associated with higher revision rates (5.28%) compared to MoP (4.28%) and CoP (3.52%) implants in the Medicare population. Both complication and revision rates were comparable to CoC implants.
Keywords:total hip arthroplasty  meta-on-metal  metal-on-polyethylene  ceramic-on-ceramic  ceramic-on-polyethylene  Medicare
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏