首页 | 本学科首页   官方微博 | 高级检索  
     


Up to 18-Year Follow-Up Wear Analysis of a First-Generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty
Authors:James E. Feng  David Novikov  Kevin Chen  Kelvin Kim  Jared Bookman  Afshin A. Anoushiravani  Matthew Hamula  Nimrod Snir  Ran Schwarzkopf
Affiliation:1. Division of Adult Reconstructive Surgery, NYU Langone Orthopedics, NYU Langone Health, New York, NY;2. Division of Orthopaedic Surgery, Albany Medical Center, Albany, New York, NY;3. Division of Adult Reconstruction, Department of Orthopaedics, Sorasky Medical Center, Tel-Aviv University, Tel Aviv, Israel
Abstract:

Background

The advent of highly cross-linked polyethylene (HCLPE) has significantly improved total hip arthroplasty survivorship. HCLPE has been shown to improve wear properties in midterm outcomes when compared to traditional polyethylene liners; however, there is a paucity of studies evaluating long-term outcomes. In addition, there is concern that wear rates may accelerate as the implant ages. Thus, the aims of this study are to report on the longest-to-date follow-up of a specific first-generation HCLPE liner and to determine whether there is a change in the annual wear rate over time.

Methods

Forty hips in 38 patients which were previously reported on in a midterm study were included in this long-term follow-up study. Patients in this cohort all received total hip arthroplasty between March 1999 and August 2004 using the Crossfire HCLPE liner. Annual wear rates (mm/y) were calculated for this cohort. Patients were contacted and asked about complications or revision procedures they may have had since the index procedure.

Results

Clinical follow-up averaged 12.9 years with a range of 7-18 years. The average follow-up duration was 12.5 years with a range of 10-17 years. Linear wear was found to be 0.056 ± 0.036 mm/y. Osteolysis was not observed in any of the patients with greater than 10-year radiographic follow-up. Furthermore, only 1 patient required revision surgery following a mechanical fall.

Conclusion

Our study demonstrates the long-term wear rates associated with HCLPE liners continue to match rates published in midterm studies. Previously, we have reported that this cohort had an average annual wear rate of 0.05 mm/y over 10 years. This most recent report demonstrates a similar wear rate with up to 18-year follow-up.
Keywords:highly cross-linked polyethylene  total hip arthroplasty  wear rates  osteolysis  acetabular liner
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号