Validation of the dual radiotracer method for quantitative In-111 platelet scintigraphy |
| |
Authors: | W J Powers K T Hopkins M J Welch |
| |
Affiliation: | Edward Mallinckrodt Institute of Radiology and Department of Neurology Washington University School of Medicine 510 South Kingshighway Blvd. St. Louis, Missouri 63110 USA |
| |
Abstract: | We have developed a simple in vivo scintigraphic technique that permits accurate quantitative comparison of intravascular platelet deposition in blood vessels of similar size. Regional count information from scintigraphic images of In-111 platelets and Tc-99m red blood cells is used to determine the ratio of radioactivity in the thrombus relative to that in the blood. Multiplication of this ratio by the percent injected dose (%ID) of In-111 per ml of blood yields a quantity (%ID index) that is directly proportional to the %ID of In-111 platelets in the thrombus as measured in vitro (r = .879, p less than 0.005). This technique is well suited for quantitative studies of scintigraphically detectable platelet deposition in both animals and in man. |
| |
Keywords: | thrombosis blood platelets indium platelets radionuclide imaging cardiovascular models |
本文献已被 ScienceDirect PubMed 等数据库收录! |