首页 | 本学科首页   官方微博 | 高级检索  
     


Rapid data acquisition protocol in ECG-gated myocardial perfusion SPECT with Tc-99m-tetrofosmin
Authors:Shin-ichiro Kumita  Tatsuo Kumazaki  Keiichi Cho  Sunao Mizumura  Tetsuji Kijima  Makiko Ishihara  Hidenobu Nakajo  Junko Sano  Yumiko Tada  Shunta Sakai  Yoshiki Kusama  Kazuo Munakata
Affiliation:1. Department of Radiology, Nippon Medical School, 1-1-5 Sendagi, Bunkyoku, 113-8603, Tokyo, Japan
2. First Department of Internal Medicine, Nippon Medical School, Japan
Abstract:
Into 25 patients with heart disorders,99mTc-tetrofosmin 555–740 MBq was injected intravenously at rest. After 40 minutes, ECG-gated myocardial perfusion SPECT was performed with a two detector gamma camera VERTEX (ADAC), setting up two detectors to form a 90-degree angle. Sixteen frames per R-R interval were acquired during a 180° rotation from the RAO 45° to the LPO 45°. A pair of data sets with standard (SDA) and rapid data acquisition (RDA) protocols was collected. In an SDA protocol, SPECT imaging was performed for 50 sec per step in 5° angular steps (total acquisition time; 15 minutes). An RDA protocol was conducted with 12 sec per step, 6° angular steps (acquisition time, 3 minutes). LVEF (%) and LVEDVml quantitated automatically with a QGS program showed excellent correlations between two protocols with correlation coefficients of 0.980 (p < 0.01) and 0.983 (p < 0.01), respectively. Subsequently visual assessment of regional wall motion based on a four-point grading system was carried out with a 3-D cine LV display. High complete agreement was gained with 158 (90.3%) out of total 175 segments, so that assessment of the global and regional LV function with the RDA protocol demonstrated high reliability and feasibility.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号