Infusion versus bolus of an ultrasound contrast agent: in vivo dose-response measurements of BR1 |
| |
Authors: | Correas J M Burns P N Lai X Qi X |
| |
Affiliation: | Department of Radiology, University Hospital of Necker, Paris, France. |
| |
Abstract: | RATIONALE AND OBJECTIVES: To determine the efficacy of an ultrasound contrast agent infusion using Doppler intensitometry estimation of backscatter enhancement in blood. METHODS: Multiple intravenous injections of BR1 (SonoVue) were performed in chronic dog studies, using bolus (0.05-2 mL) and infusion (3-40 mL/h during 6 minutes) administration. The pulsed Doppler signal from the femoral artery was recorded and analyzed for mean Doppler power and integrated fractional enhancement. RESULTS: For bolus injection, time-intensity curves exhibited a rapid first pass (peak 30 dB for 0.45 mL) followed by a slower washout. Integrated fractional enhancement exhibited a linear relation with the dose (R2 = 0.99). For infusion administration, peak enhancement increased with the infusion rate from 8 to 22 dB. At rates exceeding 30 mL/h, the enhancement was stable with a plateau-like pattern. CONCLUSIONS: Infusion of BR1 is easily achieved and allows the duration of enhancement to be increased as long as desired. Stable enhancement is obtained for rates greater than 30 mL/h. |
| |