首页 | 本学科首页   官方微博 | 高级检索  
检索        


Image-directed percutaneous biopsies with a biopsy gun
Authors:Parker  SH; Hopper  KD; Yakes  WF; Gibson  MD; Ownbey  JL; Carter  TE
Institution:Department of Radiology, Fitzsimons Army Medical Center, Aurora, Colo.
Abstract:Core tissue for histologic study is believed by many pathologists to be more diagnostic than material from needle aspiration. Recently, a biopsy "gun" has been introduced, which simplifies core biopsies. With this device, 182 biopsies of multiple anatomic sites were performed with ultrasonic, computed tomographic, and fluoroscopic guidance and 18-gauge needles. High-quality histopathologic specimens were obtained in 177 of the biopsies, and diagnostic target tissue was obtained in 167. Only three significant complications occurred: one bleeding complication that required transfusion and two cases of pneumothorax that necessitated placement of chest tubes. The biopsy gun eliminated the disjointed movements of conventional "skinny" needle biopsies, and none of the samples demonstrated significant "crush" artifact or obscuring blood, problems that are commonly associated with manual biopsy techniques. Patient discomfort was decreased with this system compared with that of manual biopsies, and the total procedure time was reduced. Because of these distinct advantages, the authors now use the biopsy gun exclusively for all percutaneous biopsies and recommend that other institutions consider the use of this biopsy method.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载免费的PDF全文