Quality control guidelines for testing gram-negative control strains with polymyxin B and colistin (polymyxin E) by standardized methods |
| |
Authors: | Jones Ronald N,Anderegg Tamara R,Swenson Jana M Quality Control Working Group |
| |
Affiliation: | The JONES Group/JMI Laboratories, Inc., 345 Beaver Kreek Centre, Suite A, North Liberty, IA 52317, USA. ronald-jones@jmilabs.com |
| |
Abstract: | An eight-laboratory study addressed the urgent need for quality control (QC) ranges for susceptibility determination when testing colistin (polymyxin E) and polymyxin B, two polycationic peptide antimicrobial agents, against multidrug-resistant gram-negative bacilli. For Escherichia coli ATCC 25922l, the QC ranges were as follows: for colistin, 0.25 to 1 microg/ml (11 to 17 mm), and for polymyxin B, 0.25 to 2 microg/ml (13 to 19 mm). For Pseudomonas aeruginosa ATCC 27853, the QC ranges were as follows: for colistin, 0.25 to 2 microg/ml (11 to 17 mm), and for polymyxin B, 0.25 to 2 mug/ml (14 to 18 mm). More than 97% of all reported QC results were within these proposed ranges. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|