首页 | 本学科首页   官方微博 | 高级检索  
检索        

诊断TracerlabFX—FDG合成18-FDG低产率故障实例
引用本文:梁秀艳,胡玉民,王文志,于丽娟.诊断TracerlabFX—FDG合成18-FDG低产率故障实例[J].医疗设备信息,2007,22(11):110-111.
作者姓名:梁秀艳  胡玉民  王文志  于丽娟
作者单位:哈尔滨医科大学附属肿瘤医院PET/CT中心,黑龙江哈尔滨150081
摘    要:详细分析了参数记录数据在Tracerlab FX-FDG上的合成过程,以及造成合成故障的确切原因,从而得出正确的结论,即:对液体的添加过程是否正常,数据记录可提供准确信息,是解决问题可依赖的分析手段。

关 键 词:自动合成率  18F—FDG  PET/CT
文章编号:1007-7510(2007)11-0110-02
收稿时间:2007-03-23
修稿时间:2007-04-24

Example of Analysis on Low Yield of 18F-FDG in Tracerlab FX-FDG
LIANG Xiu-yan, HU Yu-min, WANG Wen-zhi, YU Li-juan.Example of Analysis on Low Yield of 18F-FDG in Tracerlab FX-FDG[J].Information of Medical Equipment,2007,22(11):110-111.
Authors:LIANG Xiu-yan  HU Yu-min  WANG Wen-zhi  YU Li-juan
Abstract:Analyzes in detail the recorded data of synthesis process on Tracerhb FX-FDG and gets the factors of low yield of 18F-FDG. Obtains the correct analysis result finally and concludes if the liguid addimg process is normal and that the recoded data could be a reasonable and dependable analysis means.
Keywords:auto-synthesis  18F-FDG  PET/CT