GZP型60Co源剂量学参数的蒙特卡洛模拟
引用本文:王先良,袁珂,唐斌,康盛伟,黎杰,肖明勇,李晓兰,李林涛,王培.GZP型60Co源剂量学参数的蒙特卡洛模拟[J].中华放射肿瘤学杂志,2016,25(5):489-495.
作者姓名:王先良  袁珂  唐斌  康盛伟  黎杰  肖明勇  李晓兰  李林涛  王培
作者单位:610041 成都,四川省肿瘤医院放疗科
基金项目:四川省科技支撑计划项目“女性肿瘤疾病的综合防治研究”(2014SZ0001)Science and Technology Support Program of Sichuan Province " Comprehensive prevention and treatment of female tumor disease"(2014SZ0001)
摘    要:目的 GZP型60Co源高剂量率后装机在临床中已有应用,模拟计算GZP型60Co源的剂量学参数。方法 使用EGSnrc蒙特卡洛软件模拟计算已知的BEBIG 60Co源(Co0.A86)剂量学参数,与其结果进行对比,验证方法的可行性。对GZP型高剂量率后装机60Co源进行建模,用同样方法模拟计算GZP型60Co源剂量学参数。结果 对BEBIG60Co源,结果与标准数据吻合很好,单位活度空气比释动能强度SK/A相差0.2%,剂量率常数Λ相差1.0%,径向剂量函数gL (r)和各向异性函数F (r,θ))曲线吻合。计算得到的GZP型60Co源(1、2)号通道的SK/A和Λ分别是3.011×10-7cGycm2h-1Bq-1和1.118cGyh-1U-1, GZP (3)号通道60Co源的SK/A和Λ分别是3.002×10-7cGycm2h-1Bq-1和1.110cGyh-1U-1,gL (r)、F (r,θ)和水模中单位空气比释动能强度的剂量率参照AAPM推荐列出。

关 键 词:近距离疗法    60Co源    剂量学参数    蒙特卡洛模拟  

A Monte Carlo-based dosimetric study of the GZP 60Co source
Wang Xianliang,Yuan Ke,Tang Bin,Kang Shengwei,Li Jie,Xiao Mingyong,Li Xiaolan,Li Lintao,Wang Pei.A Monte Carlo-based dosimetric study of the GZP 60Co source[J].Chinese Journal of Radiation Oncology,2016,25(5):489-495.
Authors:Wang Xianliang  Yuan Ke  Tang Bin  Kang Shengwei  Li Jie  Xiao Mingyong  Li Xiaolan  Li Lintao  Wang Pei
Institution:Department of Radiotherapy,Sichuan Cancer Hospital,Chengdu 610041,China
Abstract:Objective To simulate and calculate the dosimetric parameters of the GZP 60Co source that has been clinically used in high-dose-rate brachytherapy.Methods The EGSnrc Monte Carlo software was used to simulate and calculate the dosimetric parameters of a well known BEBIG 60Co source (Co0.A86).The results were compared with the actual parameters to verify the feasibility of this method.A Monte Carlo model of the GZP 60Co source for high-dose-rate brachytherapy was established to simulate and calculate its dosimetric parameters in the same way.Results For the BEBIG 60Co source,the resuhs were well accorded with the standard.The air-kerma strength per unit activity (SK/A) and dose rate constant (∧)deviated from the standard by 0.2% and 1.0%,respectively.The curves of the radial dose function gL(r) and the anisotropy function F (r,θ) fit well.For the GZP 60Co source,the SK/A and ∧values were calculated as 3.011 × 10-7 cGycm2h-1Bq-1 and 1.118 cGyh-1 U-1 in channel l&2 and 3.002× 10-7 cGycm2h-1 Bq-1 and 1.110 cGyh-1U-1 in channel 3.The gL(r),F (r,θ),and dose rate per unit air-kerma strength in a liquid water phantom were listed as the American Association of Physicists in Medicine recommended.Conclusions The results can be used in planning system and quality control for the GZP 60Co source.
Keywords:Brachytherapy  60Co source  Dosimetric parameters  Monte Carlo simulation