Detection of immunoglobulins G and M to rubella virus by time-resolved immunofluorometry. |
| |
Authors: | P Shankaran E Reichstein M J Khosravi E P Diamandis |
| |
Affiliation: | CyberFluor Inc., Toronto, Ontario, Canada. |
| |
Abstract: | We describe new methods for the detection of immunoglobulin G (IgG) and IgM rubella-specific antibodies in serum. The IgG assay was based on a solid-phase rubella antigen immobilization approach, and the IgM assay was based on the IgM capture assay principle. Both assays used biotinylated antibodies (anti-human IgG and antirubella monoclonal antibody, respectively). The tracer system was based on streptavidin labeled with a fluorescent europium chelate. The final measurements were done by using time-resolved fluorescence. Both assays were thoroughly evaluated with clinical samples and compared successfully with established techniques. We anticipate that these assays are suitable for routine clinical use. |
| |
Keywords: | |
|
|