Evaluation of BacT/Alert 3D liquid culture system for recovery of mycobacteria from clinical specimens using sodium dodecyl (lauryl) sulfate-NaOH decontamination |
| |
Authors: | Carricajo A Fonsale N Vautrin A C Aubert G |
| |
Affiliation: | Department of Bacteriology, CHU Bellevue Hospital, Saint-Etienne, France. anne.carricajo@uni-st-etienne.fr |
| |
Abstract: | A total of 52 mycobacterial isolates were recovered from 1,197 clinical specimens decontaminated by a sodium dodecyl (lauryl) sulfate (SDS)-NaOH protocol. Of these, 94% were recovered with the BacT/Alert 3D system (Organon Teknika, Durham, N.C.) and 79% were recovered on L?wenstein-Jensen (LJ) medium. Mean times to detection of organisms of the Mycobacterium tuberculosis complex (n = 47) were 22.8 days with LJ medium and 16.2 days with the system. The BacT/Alert 3D system is a rapid and efficient detection system which can be used with an SDS-NaOH decontamination procedure. |
| |