Effect of radiographic techniques (kVp and mAs) on image quality and patient doses in digital subtraction angiography
Authors:Gkanatsios Nikolaos A  Huda Walter  Peters Keith R
Affiliation:Department of Radiology, University of Florida, Gainesville 32610, USA.
Abstract:We investigated how varying the x-ray tube voltage and image receptor input exposure affected image quality and patient radiation doses in interventional neuroradiologic imaging. Digital subtraction angiography (DSA) images were obtained of a phantom with 1 mm diameter vessels containing iodine at concentrations between 4.5 and 50 mg/cc. The detection threshold concentration of iodine was determined by inspecting DSA images obtained at a range of x-ray tube voltages and input exposure levels. Surface doses were obtained from measured x-ray tube output data, and corresponding values of energy imparted were determined using the exposure-area product incident on the phantom. In one series of experiments, the air kerma at the image intensifier (X) was varied between 0.44 microGy per frame and 8.8 microGy per frame at a constant x-ray tube voltage of 70 kVp. In a second series of experiments, the tube voltage was varied between 50 and 100 kVp, and the mAs adjusted to maintain a constant exposure level at the input of the image intensifier. At a constant x-ray tube voltage, the surface dose and energy imparted were directly proportional to the input exposure per frame used to acquire the DSA images. On our DSA system operated below 2.2 microGy per frame, the threshold iodine concentration was found to be proportional to X(-0.57), which is in reasonable agreement with the theoretical prediction for a quantum noise limited imaging system. Above 2.2 microGy per frame, however, the threshold iodine concentration was proportional to X(-0.26), indicating that increasing the input exposure above this value will only achieve modest improvements in image quality. At a constant image intensifier input exposure level, increasing the x-ray tube voltage from 50 kVp to 100 kVp reduced the surface dose by a factor of 6.1, and the energy imparted by a factor of 3.5. The detection threshold iodine concentration was found to be proportional to kVp(n), where n was 2.1 at 1.1 microGy per frame, and 1.6 at 3.9 microGy per frame. For clinical situations that can be modeled by a uniform phantom, reducing the x-ray tube voltage rather than increasing the exposure level would best achieve improvements on our DSA imaging system performance.
Keywords:
本文献已被 PubMed 等数据库收录!
正在获取引用信息,请稍候...
正在获取相似文献,请稍候...
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号