Formation of thrombin-antithrombin III complex using polyamide and hemophan dialyzers. |
| |
Authors: | G Schultze S Hollmann P Sinah |
| |
Affiliation: | Humboldt-Krankenhaus, Berlin, Germany. |
| |
Abstract: | The recently developed ELISA for the thrombin-antithrombin III complex (TAT) is a sensitive, specific, and simplified means of detecting intravascular coagulation. For further evaluation of the thrombogenicity of a polyamide (P) and a Hemophan (H) hollow-fibre dialyzer a cross-over study was done in ten stable patients on maintenance hemodialysis. At the same doses of heparin (mean bolus of 30 U/kg bw and maintenance doses of 86 U/kg bw), thrombin time and partial thromboplastin time were significantly lower using H. At the end of dialysis TAT was significantly higher in H (mean +/- SEM before HD 3.57 +/- .56, at 240 min 14.9 +/- 6.5 ng/ml, p less than 0.05, Wilcoxon-test) than in P (before HD 4.36 +/- .98, at 240 min 8.95 +/- 3.0 ng/ml, p less than 0.05 H 240 vs. P 240, Wilcoxon-test). Visible clotting was more pronounced in the H filter. Among other favourable features of blood compatibility the polyamide/polyvinylpyrrolidone copolymer with a hydrophilic/hydrophobic microdomain structure has less thrombogenicity. The modified cellulosic membrane H has advantages in complement activation and leukocyte depression, but thrombogenicity seems less favourable since the incorporated diethyl-amino-ethyl groups with their positive charge bind and inactivate negatively charged heparin. |
| |