首页 | 本学科首页   官方微博 | 高级检索  
     


Fixation Strength of Polyetheretherketone Sheath-and-Bullet Device for Soft Tissue Repair in the Foot and Ankle
Authors:Jay Christensen  Brian Fischer  Michael Nute  Robert Rizza
Abstract:
Tendon transfers are often performed in the foot and ankle. Recently, interference screws have been a popular choice owing to their ease of use and fixation strength. Considering the benefits, one disadvantage of such devices is laceration of the soft tissues by the implant threads during placement that potentially weaken the structural integrity of the grafts. A shape memory polyetheretherketone bullet-in-sheath tenodesis device uses circumferential compression, eliminating potential damage from thread rotation and maintaining the soft tissue orientation of the graft. The aim of this study was to determine the pullout strength and failure mode for this device in both a synthetic bone analogue and porcine bone models. Thirteen mature bovine extensor tendons were secured into ten 4.0?×?4.0?×?4.0-cm cubes of 15-pound per cubic foot solid rigid polyurethane foam bone analogue models or 3 porcine femoral condyles using the 5?×?20-mm polyetheretherketone soft tissue anchor. The bullet-in-sheath device demonstrated a mean pullout of 280.84 N in the bone analog models and 419.47 N in the porcine bone models. (p?=?.001). The bullet-in-sheath design preserved the integrity of the tendon graft, and none of the implants dislodged from their original position.
Keywords:5  bone model  fixation strength  interference device  soft tissue anchor  tendon transfer
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏