High Stroke Volume Para-aortic Counterpulsation Device versus Centrifugal Pump in Cardiogenic Shock: Experimental Study
Authors:Serafim N. Nanas  John N. Nanas  Christos E. Charitos  Alexandros Gougoulakis  Konstantinos Makaritsis  John Chatzigeorgiou  Kiriaki Moussoutzani  Maria I. Anastasiou-Nana  Spyridon D. Moulopoulos
Affiliation:(1) Department of Clinical Therapeutics, National and Kapodistrian University of Athens, Vas. Sofias 80 & K. Lourou, Athens 11518, Greece, GR
Abstract:p < 0.001), the tension time index (712 ± 381 versus 1333 ± 694, p < 0.01), and the double product (5629 ± 2574 versus 7440 ± 3294, p < 0.01) were significantly lower during assistance with the PACD than with the CBP. It was concluded that PACD is at least as effective as CBP for restoring hemodynamic status during acute experimental cardiogenic shock. Moreover, the PACD unloads the left ventricle more effectively than CBP, making it suitable for left ventricular mechanical support in cases with reversible myocardial damage.