Reduction of RV pacing by continuous optimization of the AV interval |
| |
Authors: | Milasinovic Goran,Sperzel Johannes,Smith Timothy W,Mead Hardwin,Brandt Johan,Haisty Wesley K,Bailey J Russell,Roelke Marc,Simonson Jay,Gerritse Bart,Englund Jennifer,Compton Steven J Worldwide EnPulse Investigators |
| |
Affiliation: | Pacemaker Center, Clinical Center of Serbia, Belgrade, Serbia and Montenegro. goran_milas@yahoo.com |
| |
Abstract: | BACKGROUND: In patients requiring permanent pacing, preservation of intrinsic ventricular activation is preferred whenever possible. The Search AV+ (SAV+) algorithm in Medtronic EnPulsetrade mark dual-chamber pacemakers can increase atrioventricular (AV) intervals to 320 ms in patients with intact or intermittent AV conduction. This prospective, multicenter study compared the percentage of ventricular pacing with and without AV interval extension. METHODS: Among 197 patients enrolled in the study, the percentage of ventricular-paced beats was evaluated via device diagnostics at the 1-month follow-up. Patient cohorts were defined by clinician assessment of conduction via a 1:1 AV conduction test at the 2-week follow-up. The observed percentage of ventricular pacing with SAV + ON and the predicted percentage of ventricular pacing with SAV + OFF were determined from the SAV + histogram data for the period between the 2-week and 1-month follow-up visits. RESULTS: Of 197 patients, 110 (55.8%) had intact 1:1 AV conduction, of which 109 had 1-month data. SAV + remained ON in 99/109 patients; 10 patients had intrinsic A-V conduction intervals beyond SAV + nominal and therefore SAV + disabled. The mean percentage of ventricular pacing in the 109 patients was SAV+ ON = 23.1% (median 3.7%) versus SAV + OFF = 97.2% (median 99.7%). In 87 patients without 1:1 AV conduction, SAV + was programmed OFF in 6, automatically disabled in 52, and remained ON in 29. In 8 of these patients, 80-100% reduction in ventricular pacing was observed with SAV + ON. CONCLUSION: The Search AV+ algorithm in the EnPulse pacemaker effectively promotes intrinsic ventricular activation and substantially reduces unnecessary ventricular pacing. |
| |
Keywords: | search AV+ dual-chamber pacing intrinsic ventricular activation unnecessary ventricular pacing |
本文献已被 PubMed 等数据库收录! |
|