The influence of carbon dioxide,bicarbonate and other buffers on the potential of antimony microelectrodes
Authors:P. Quehenberger  S. David
Affiliation:(1) Institut für Physiologie und Balneologie der Universität, A-6010 Innsbruck, Austria
Abstract:
Summary Antimony microelectrodes were calibrated at 37° C in phosphate buffers, in different bicarbonate solutions at various CO2-partial pressures and in buffers like TRIS, TES, MES and malonic acid. By use of the latter buffers (with exception of malonic acid) the most reliable calibration curves were obtained (ldquonormal valuesrdquo). The usual calibration in 67 mmol/l standard phosphate buffers turned out to be unacceptable bacause the obtained mV-values were too high (negative) in comparison to all other buffers. Different calibration curves resulted from the use of pure bicarbonate solutions whether the pH-values were changed by variation ofpCO2 or of the bicarbonate concentration. Low bicarbonate concentrations in combination with lowpCO2 gave mV-values which were too low relative to the other buffers. Both the increase ofpCO2 as well as of the bicarbonate concentration caused a shift of the potential of the antimony electrodes toward ldquonormal valuesrdquo. In solutions containing other buffers the influence of bicarbonate andpCO2 became negligible with increasing buffer concentration.Decreasing oxygen partial pressure was found to cause an increase of the potential of the antimony electrodes.The influence of liquid junction potentials at the reference electrode is discussed.
Keywords:Atimony  Microelectrode  Hydrogen ion concentration  Phosphate  Bicarbonate  Carbon dioxide
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号