Left ventricular unloading and concomitant total cardiac output increase by the use of percutaneous Impella Recover LP 2.5 assist device during high-risk coronary intervention.
Authors:Marco Valgimigli  Paul Steendijk  George Sianos  Emile Onderwater  Patrick W Serruys
Affiliation:Department of Interventional Cardiology, Thoraxcenter, Erasmus MC, Rotterdam, The Netherlands.
Abstract:
A number of techniques have been proposed for circulatory support during high-risk percutaneous coronary interventions (PCI), but no single approach has achieved wide acceptance so far. We report on a patient with severe left ventricular (LV) impairment who underwent a PCI with the use of a new left ventricular assist device, the Impella Recover LP 2.5 system. The effects on global cardiac output were determined by thermodilution (TD) and LV pressure-volume loops obtained by conductance catheter. The activation of the pump resulted in a rapid and sustained unloading effect of the LV. At the same time, the continuous expulsion of blood into ascending aorta throughout the cardiac cycle produced by the pump resulted in an increase of systemic overall CO, measured by the TD technique, of 1.43 L/min. The procedure was uncomplicated and the patient remained uneventful at follow-up. Our single experience gives new input for future trials to assess the effect of the Impella Recover LP 2.5 assist device on outcome in this subset of patients.
Keywords:left ventricular assist device  high‐risk percutaneous coronary intervention  pressure‐volume loop