Surgical device closure a new technique of ASD closure |
| |
Authors: | Mahadev Dixit Mohan Gan Anuradha Dubey Narendra Nishanimath Pradeep Kizakke Kuttikot Raghunath Mohapatra Darshan Devikere Sharabhalingswamy Raghav Johari Bhaskar Venkatakrishnaiah Bhagyalakshmi Aruneshwari Dayal |
| |
Affiliation: | (1) Departments of Cardiothoracic Surgery, Cardiology and Cardiac Anaesthesia, JN Medical College and KLES Hospital and MRC Belgaum, Nehru Nagar, Belgaum, 590010, Karnataka, India |
| |
Abstract: | Background Conventional approach of atrial septal defect (ASD) closure with cardio pulmonary bypass using mid sternotomy, minimally invasive or endoscopic technique is time tested. We decided to use custom made device with direct minimally invasive approach without cardio pulmonary bypass. Percutaneous transfemoral route using custom made device is a well established procedure performed by interventional cardiologist with occasional trauma and vessels. Method We performed the procedure in 3 patients of secundum ASD deemed adequate for device closure. We used a mini Right anterior thoracotomy approach using a double umbrella device which was implanted through direct Right Atrial puncture. Results In two patients we were successful in deploying the device. The proedure lasted 30 minutes with small infra Mammary scar and the post operative period was uneventful. In the third case where our attempt at surgical device closure failed the inferior margin was only one mm. The optimal size device kept slipping into the right atrium when the guide-wire was tugged after final deployment. The placement of an oversize device distorted the mitral valve, may causing regurgitation. Conclusion We think that this is a simple and safe technique of secundum ASD closure without cardio pulmonary bypass. |
| |
Keywords: | Atrial septal defect Echocardiography Thoracotomy |
本文献已被 SpringerLink 等数据库收录! |