首页 | 本学科首页   官方微博 | 高级检索  
     


The Parker Flex-Tip Tube versus a Standard Tube for Fiberoptic Orotracheal Intubation: A Randomized Double-blind Study
Authors:Kristensen   Michael S. M.D.
Affiliation:Kristensen, Michael S. M.D.*
Abstract:
Background: During fiberoptic tracheal intubation, passage of the fiberscope itself to the trachea is often fairly easy, but passage of the tube into the trachea may be difficult or even impossible. A new type of disposable endotracheal tube, the Parker Flex-Tip tube, has a tip that reduces the gap between the fiberscope and the inside of the tube. Thus, theoretically, a smaller risk of impinging on laryngeal structures during insertion in trachea is expected.

Methods: Eighty patients scheduled for elective anesthesia using orotracheal intubation were randomized to either a Parker Flex-Tip tube or a standard (Portex) 7.5-mm-ID endotracheal tube. Blinding was obtained by having the tube premounted on the fiberscope (Olympus LF-1; diameter of fiberscope = 4 mm) and thereafter covered with a black opaque plastic bag. Difficulty in placing the tube was scored using an objective standardized grading system.

Results: Seventy-six patients completed the study. The use of the Parker Flex-Tip tube reduced the incidence of need for repositioning of the tube during insertion into trachea from 89% to 29% (P < 0.0001) when compared to the standard tube. The median time for passage of the tube into the trachea was reduced from 20 s to 7.5 s (P < 0.0001).

Keywords:
点击此处可从《Anesthesiology》浏览原始摘要信息
设为首页 | 免责声明 | 关于勤云 | 加入收藏