首页 | 本学科首页   官方微博 | 高级检索  
     

用双试剂与日立7170A分析仪测定血清无机磷
引用本文:王断贵 蒋洪敏. 用双试剂与日立7170A分析仪测定血清无机磷[J]. 湖南医科大学学报, 1997, 22(5): 455-458
作者姓名:王断贵 蒋洪敏
摘    要:用日立7170A分析仪以双试剂测定血清无机磷,并观察异常蛋白引起的干扰。该法线性范围可达5mmol.L^-1,准确度和精密度系数均在95%以上,回收率在98.2% ̄99.5%之间,脂浊血清及黄疸表对本法不产生干扰。双试剂与多发性骨髓瘤病人血清(含异常蛋白)不产生沉淀,是一种理想的血清无机磷测定法。

关 键 词:磷酸盐类 无机磷 双试剂 血清

Determination of inorganic phosphate in serum by double reagents with Hitachi 7170 A analyzer]
J Wang,H Jiang. Determination of inorganic phosphate in serum by double reagents with Hitachi 7170 A analyzer][J]. Bulletin of Hunan Medical University, 1997, 22(5): 455-458
Authors:J Wang  H Jiang
Affiliation:Department of Clinical Laboratory, Second Affiliated Hospital, Hunan Medical University, Changsha.
Abstract:We propose a method for determination of inorganic phosphate in serum using double reagents with Hitachi 7170 A analyzer. The standard curve for phosphate was linear from 1.0-5.0 mmol.L-1. Both coefficient of accuracy and precision were above 95%. The recovery rate ranged from 98.2%-99.5%. Jaundice Sample and lipohemia serum were found to be interference--free. This method can eliminate paraprotein interference, it was an ideal method for determination of inorganic phosphate in serum.
Keywords:
本文献已被 维普 等数据库收录!