采用门控技术克服心脏搏动对肝脏DWI影响的相关研究
引用本文:李勇坚,向之明,黄芳,梁倩雯,李建安.采用门控技术克服心脏搏动对肝脏DWI影响的相关研究[J].中国CT和MRI杂志,2017(8):91-93.
作者姓名:李勇坚  向之明  黄芳  梁倩雯  李建安
作者单位:1. 广东省广州市番禺区南村医院 广东 广州 511422;2. 广东省广州市番禺区中心医院 广东 广州 511400;3. 广东省第二中医院 广东 广州 510095
基金项目:广州市科技计划项目;项目(2015100010002)
摘    要:目的本文旨在探讨心电门控技术克服心脏搏动对肝脏DWI的影响效果。方法选择志愿者41例,均行无门控和心电门控肝脏DWI检查,心电门控组为实验组,无门控组为对照组,b值选取0和500s/mm2,分别采用不同的延迟触发时间触发扫描。第一次肝脏DWI扫描为10点和19点,采集2次的ADC值,由2名测量者分别测量,测完第一次数据后,间隔15天再进行第二次测量,总结分析ADC值,得出结论。结果对照组肝左叶ADC值明显高于肝右叶;实验组肝左叶ADC值明显低于肝右叶;Bland-Altman分析显示,该方法可重复率高;延迟触发时间为500ms时,可重复率高。结论心电门控技术能有效降低ADC值,提高肝脏DWI扫描图像质量,可重复性高,最佳触发时间为500ms。采用心电门控技术克服心脏搏动效果比较理想,适宜推广使用。

关 键 词:门控技术  克服心脏搏动  肝脏DWI  影响  研究

The Effect of the Liver DWI by the use of ECG-gating Technology to Overcome the Heart Beat
Abstract:Objective The purpose of this paper is to explore effect of the ECG-gating technology of liver DWI effects can overcome heart throb.Methods Select volunteers,41 cases were unable to control and switch control liver DWI examination,can the heart switch control group as the experimental group,unable to control group to the control group, s/b value selection of 0 and 500 mm squared,adopt different triggering time delay trigger scanning respectively. First liver DWI scans for 10 points and 19 points,can collect two ADC values,measured by two measurement,measurement is 1 after measuring the data for the first time, interval of 15 days and then to the second measurement, summarizes the analysis of ADC values, draw the conclusion.Results The control group of the left hepatic lobe ADC values were significantly higher than those of the right liver lobe. The experimental group of the left hepatic lobe ADC values significantly below the right liver lobe. Bland-Altman analysis shows that this method can repeat rate is high. When trigger delay time of 500 ms, can repeat rate is high.Conclusion Heart switch keying technology can effectively reduce the ADC values,improve the quality of liver DWI scans can, high repeatability, best to trigger in the 500 ms.The heart switch control technology to overcome heart beat effect is more ideal, suitable for promotion use.
Keywords:The Door Control Technology  Overcome the Heart Beat  The Liver DWI  Effects  Research