Particle size and morphology of UHMWPE wear debris in failed total knee arthroplasties--a comparison between mobile bearing and fixed bearing knees.
Authors:Chun-Hsiung Huang  Fang-Yuan Ho  Hon-Ming Ma  Chan-Tsung Yang  Jiann-Jong Liau  Hung-Chan Kao  Tai-Horng Young  Cheng-Kung Cheng
Institution:Department of Orthopaedic Surgery, Mackay Memorial Hospital, Taipei, Taiwan. chhuang@ms2.mmh.org.tw
Abstract:Osteolysis induced by ultrahigh molecular weight polyethylene wear debris has been recognized as the major cause of long-term failure in total joint arthroplasties. In a previous study, the prevalence of intraoperatively identified osteolysis during primary revision surgery was much higher in mobile bearing knee replacements (47%) than in fixed bearing knee replacements (13%). We postulated that mobile bearing knee implants tend to produce smaller sized particles. In our current study, we compared the particle size and morphology of polyethylene wear debris between failed mobile bearing and fixed bearing knees. Tissue specimens from interfacial and lytic regions were extracted during revision surgery of 10 mobile bearing knees (all of the low contact stress (LCS) design) and 17 fixed bearing knees (10 of the porous-coated anatomic (PCA) and 7 of the Miller/Galante design). Polyethylene particles were isolated from the tissue specimens and examined using both scanning electron microscopy and light-scattering analyses. The LCS mobile bearing knees produced smaller particulate debris (mean equivalent spherical diameter: 0.58 microm in LCS, 1.17 microm in PCA and 5.23 microm in M/G) and more granular debris (mean value: 93% in LCS, 77% in PCA and 15% in M/G).
Keywords:Polyethylene  Wear debris  Particle size  Light‐scattering  Osteolysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号