首页 | 本学科首页   官方微博 | 高级检索  
     

Sysmex XE-2100血细胞分析仪定量分析外周血幼稚粒细胞性能评价
引用本文:王晓忠,曾学辉,莫莉,周小梅. Sysmex XE-2100血细胞分析仪定量分析外周血幼稚粒细胞性能评价[J]. 江西医学检验, 2009, 0(3): 214-216
作者姓名:王晓忠  曾学辉  莫莉  周小梅
作者单位:广州中医药大学附属深圳医院检验科;深圳市福田区中医院检验科;
摘    要:目的评价Sysmex XE-2100血细胞分析仪定量分析外周血幼稚粒细胞(1G)的性能。方法选取202份XE-2100IG报警值〉200的样本.分别采用XE-2100和镜检分析IG%,评价两者之间相关性,选取5份不同IG%水平的样本进行XE-2100批内精密度(重复性)评价,分别选取8份标本在4℃和室温进行样本稳定性评价。结果XE-2100分析IG%结果与镜检结果具有良好的相关性(γ=0.8382,P〈0.001);5份不同IG%水平样本重复性的变异系数(CV)从4.93%N15.5%;XE-2100分析4℃和室温保存24小时内的样本结果稳定。结论XE-2100定量分析外周血幼稚粒细胞性能良好,基本满足临床实验室要求。

关 键 词:血细胞分析仪  幼稚粒细胞

The evaluation of quantitative measurement of immature granulocyte in peripheral blood using Sysmex XE-2100 hematology analyzer
Wang Xiaozhong,Zeng Xuehui,Mo Li,et al.. The evaluation of quantitative measurement of immature granulocyte in peripheral blood using Sysmex XE-2100 hematology analyzer[J]. Jiangxi Journal of Medical Laboratory Sciences, 2009, 0(3): 214-216
Authors:Wang Xiaozhong  Zeng Xuehui  Mo Li  et al.
Affiliation:Wang Xiaozhong,Zeng Xuehui,Mo Li,et al. Department of Laboratory Medicine,Affiliated Shenzhen Hospital of Guangzhou University of Traditional Chinese Medicine,Shenzhen 518033,China
Abstract:Objective To evaluate the performance of quantitative measurement of immature granulocyte (IG) in peripheral blood using Sysmex XE-2100 hematology analyzer. Methods 202 samples with IG alarm value 〉200 were selected to evaluate the correlation between XE-2100 count and manual morphology count, 5 samples at different levels were selected to evaluate within-run reproducibility for XE-2100 IG count, 8 specimens were selected to evaluate the sample stability at 4℃ and at room temperature. Results The correlativity analysis showed IG agreement between XE-2100 count and the manual morphology count (γ=0.8382, P〈0.001); The within-run reproducibility performed using fresh blood on 5 different level specimens showed a range from 4.93% to 15.5% CV, The percent of IG were stable at both room temperature and after refrigeration for 24 hours. Conclusions The performance of quantitative measurement of immature granulocyte in peripheral blood using Sysmex XE-2100 hematology is well, it can meet the requirements of clinical laboratories.
Keywords:Hematology analyzer  Immature granulocyte  
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号