摘    要:目的对2006~2008年参加国际临床化学联合会(IFCC)组织的参考实验室国际比对计划(RELA)的γ-谷氨酰基转移酶(GGT)测定结果进行比较和分析,为GGT参考方法的建立和应用提供借鉴。方法按照IFCC公布的酶学活性测定(37℃)参考方法的标准操作程序(SOP)测定GGT国际样本,不精密度评价按美国临床实验室标准化委员会(NCCLS)EP5-A2文件进行,准确度评价采用国际有证参考物质(ERM)。结果 2006年测定结果:样本A为(1.829±0.021)μkat/L,样本B为(3.331±0.019)μkat/L;2007年测定结果:样本A为(1.533±0.017)μkat/L,样本B为(3.840±0.035)μkat/L;2008年测定结果:样本A为(3.392±0.024)μkat/L,样本B为(3.058±0.024)μkat/L。2006~2008年GGT参考方法的不精密度分别为1.29%、0.44%、0.65%,ERM测定均值在"靶值±不确定度"[(1.90±0.04)μkat/L]范围内,初步验证了参考方法的不精密度和准确度。结论 2006~2008年RELA样本GGT测定结果均在IFCC公布的可信区间内,GGT的IFCC参考方法已经建立。

关 键 词:γ-谷氨酰基转移酶  参考值  评价研究  临床酶试验

Analysis of the RELA results of gamma-glutamyltransferase reference method
CHI Shan,SHEN Mo,ZHANG Man. Analysis of the RELA results of gamma-glutamyltransferase reference method[J]. Laboratory Medicine, 2011, 26(8): 540-543
Authors:CHI Shan  SHEN Mo  ZHANG Man
Affiliation:.(Center of Clinical Laboratory,Beijing Shijitan Hospital Affiliated to Capital Medicial University,Beijing 100038,China)
Abstract:Objective To compare and analyze the RELA [International Federation of Clinical Chemistry and Laboratory Medicine(IFCC) external quality assessment scheme for reference laboratories in laboratorg medicine] results of gamma-glutamyltransferase(GGT) from the IFCC 2006-2008 in order to provide the reference for the establishment and application of GGT reference method.Methods The GGT levels of RELA samples were measured according to the standard operation procedure(SOP)(37 ℃) published by IFCC.The precision was evaluated according to the EP5-A2 document from National Committee for Clinical Laboratory Standards(NCCLS),and the accuracy was evaluated by European reference materials(ERM).Results In RELA 2006,the result of sample A was(1.829±0.021) μkat/L,and the result of sample B was(3.331±0.019) μkat/L.In RELA 2007,the result of sample A was(1.533±0.017) μkat/L,and the result of sample B was(3.840±0.035) μkat/L.In RELA 2008,the result of sample A was(3.392±0.024) μkat/L,and the result of sample B was(3.058±0.024) μkat /L.The precisions of the year 2006,2007 and 2008 were 1.29%,0.44% and 0.65% respectively,and the results of ERM agreed with the certified value [target value±precision value:(1.90±0.04) μkat/L],which verified the precision and accuracy of the reference method.Conclusions All of the results of the 3 years agree to the confidence interval by IFCC.The GGT reference method has been established.
Keywords:Gamma-glutamyltransferase  Reference value  Evaluation study  Enzyme test