Dead-time correction in dynamic radionuclide studies by computer |
| |
Authors: | V. Ullmann V. Hušák L. Dubroka |
| |
Affiliation: | (1) Department of Nuclear Medicine, Regional Hospital, Vit února, 1790 Ostrava-Poruba, SSR;(2) Department of Nuclear Medicine, University Hospital, I.P. Pavlova 6, Olomouc, SSR;(3) Palacký University, I.P. Pavlova 6, Olomouc, SSR |
| |
Abstract: | Several methods for the measurement of dead time and for evaluation of the dependencies required to correct dynamic studies for dead-time losses are described. Two program variants were written to produce time-activity curves in the selected areas of interest of the dynamic studies with computer's correction of dead-time losses; these variants are part of the system of programs for complex processing of scintigraphic studies set up for a Clincom instrument. The first variant performs correction on the basis of the registered count rate in the whole image, while the second variant makes use of the additional reference source positioned on the periphery of the camera visual field. |
| |
Keywords: | |